95x Series Operating Manual - May 17, 2022
Page 37 of 155
In-Phase Resistance (Ωinp)
Quadrature current (Aqua)
Quadrature Voltage (Vqua)
Quadrature impedance (Ωqua)
Capacitance (F)
Dissipation Factor (DF)
Peak or breakdown current (Apk)
Arc current (Aarc)
• Many results are capable of having a negative polarity; a negative result is preceded by the – character. A
positive result has no preceding polarity character.
• The 95x automatically formats the displayed result to give the best meaningful resolution using a
multiplier character as needed (e.g., G, M, K, m, u, n, or p). Displayed results may have slightly limited
resolution; full resolution is available via an interface.
• Impedance and resistance results are calculated from voltage and current measurements. For near zero
and negative current measurements, the 95x displays the maximum meaningful resistance value
preceded by the > character.
REVIEWING TEST RESULTS AFTER RUNNING A TEST SEQUENCE
Immediately after running a test sequence and temporarily displaying an overall pass/fail status message the 95x
automatically initiates reviewing the detailed test results. The results of a previously run test sequence can also be
manually reviewed later by pressing the VIEW key if the READY indicator is illuminated, indicating that the test
sequence is still ready to be run.
If the test sequence failed, then the review is initiated at the first failed step, otherwise it is initiated at the first
test step in the test sequence.
The following table shows example displays of reviewed results and their meanings. See the descriptions following
the table for a description of each of the fields in the display.