95x Series Operating Manual - May 17, 2022
Page 40 of 155
Press the ENTER key.
The display now shows the test report printout configuration menu. An example of the test report
printout configuration menu is as follows
STATUS: ATTACHED
HEADING:
AUTO TEST REPORT: NO
TEST DETAIL: NORMAL
SERIAL #: NO
The user may navigate this sub-menu and change the displayed settings as required.
o STATUS. This shows the present status of the printer interface. The information shown is not
editable and will automatically update should a change occur.
NOT FITTED. Indicates that the USB interface is not fitted, so there is no printout
capability.
DISABLED. Indicates that the user has disabled the ability to perform printouts by
setting the DISABLE USB setting in the CNFG INTERFACES menu to YES.
NOT FOUND. Indicates that no valid printer has been detected as being connected to
the 95x USB port.
ATTACHED. Indicates that a valid printer is attached to the 95x USB port and can be
used for printouts.
o HEADING. This allows the user to enter a global 2
nd
heading for all printouts. This can be up to 12
characters long and can be any alpha-numeric characters. If this is blank, then the 2
nd
heading is
omitted from all printouts. Typically, this is used to allow the user to print their company name
or department name on all printouts.
o AUTO TEST REPORT. This allows the user to set whether the 95x will automatically print a test
report when a test sequence terminates (YES) or not (NO). The user can also command a printout
by pressing the PRINT button while reviewing test results independently of this setting.
o TEST DETAIL. This sets the amount of detail to be included in the printed test report.
BRIEF. The pass/fail status, reason for failure (if any), and the level and frequency are
printed for each test step having a result.
NORMAL. The pass/fail status, reason for failure (if any), the level and frequency, the
test limits and the maximum, minimum, average, and last measurements are printed for
each test step having a result.
FULL. The pass/fail status, reason for failure (if any), the level and frequency, the full
test step settings and the maximum, minimum, average, and last measurements are
printed for every test step.
o SERIAL #. If set to YES, the 95x will prompt the user for entry of a serial # when a test report is to
be printed (either manually or automatically). The serial number can be any alpha-numeric
characters; up to 12 characters can be included. This is included in the test report. If NO is
selected, then the user will not be prompted for the serial # and it will not be included in the
printed test report.
When finished, press the EXIT key to exit this sub-menu, and then press EXIT again (if no additional
configuring is to be performed) to exit the main configuration menu and save the settings.