95x Series Operating Manual - May 17, 2022
Page 40 of 155
• Press the ENTER key.
• The display now shows the test report printout configuration menu. An example of the test report
printout configuration menu is as follows –
STATUS: ATTACHED
HEADING:
AUTO TEST REPORT: NO
TEST DETAIL: NORMAL
SERIAL #: NO
• The user may navigate this sub-menu and change the displayed settings as required.
o STATUS. This shows the present status of the printer interface. The information shown is not
editable and will automatically update should a change occur.
▪ NOT FITTED. Indicates that the USB interface is not fitted, so there is no printout
capability.
▪ DISABLED. Indicates that the user has disabled the ability to perform printouts by
setting the DISABLE USB setting in the CNFG – INTERFACES menu to YES.
▪ NOT FOUND. Indicates that no valid printer has been detected as being connected to
the 95x USB port.
▪ ATTACHED. Indicates that a valid printer is attached to the 95x USB port and can be
used for printouts.
o HEADING. This allows the user to enter a global 2
nd
heading for all printouts. This can be up to 12
characters long and can be any alpha-numeric characters. If this is blank, then the 2
nd
heading is
omitted from all printouts. Typically, this is used to allow the user to print their company name
or department name on all printouts.
o AUTO TEST REPORT. This allows the user to set whether the 95x will automatically print a test
report when a test sequence terminates (YES) or not (NO). The user can also command a printout
by pressing the PRINT button while reviewing test results independently of this setting.
o TEST DETAIL. This sets the amount of detail to be included in the printed test report.
▪ BRIEF. The pass/fail status, reason for failure (if any), and the level and frequency are
printed for each test step having a result.
▪ NORMAL. The pass/fail status, reason for failure (if any), the level and frequency, the
test limits and the maximum, minimum, average, and last measurements are printed for
each test step having a result.
▪ FULL. The pass/fail status, reason for failure (if any), the level and frequency, the full
test step settings and the maximum, minimum, average, and last measurements are
printed for every test step.
o SERIAL #. If set to YES, the 95x will prompt the user for entry of a serial # when a test report is to
be printed (either manually or automatically). The serial number can be any alpha-numeric
characters; up to 12 characters can be included. This is included in the test report. If NO is
selected, then the user will not be prompted for the serial # and it will not be included in the
printed test report.
• When finished, press the EXIT key to exit this sub-menu, and then press EXIT again (if no additional
configuring is to be performed) to exit the main configuration menu and save the settings.