EmbryoScope time-lapse incubator
- 19 -
To navigate from the set point menu to the calibration menu, press the down arrow for about three
seconds. To navigate back from the calibration menu to the set point menu, press the up arrow for
about three seconds. While the menu is being changed, the display will flash: until
the change of menu is complete.
To navigate back and forth between the items available in each menu, press the left or right arrow.
For example, if the temperature is currently displayed in the set point menu and you want to
inspect the CO
2
concentration, press the right arrow once. If you afterwards want to return to the
temperature, press the left arrow once.
5.2 Locking or unlocking the incubator control panel
The control panel can be locked to ensure that no parameters are inadvertently changed. When
the control panel is locked, you cannot change the set point or calibrate the parameters; neither
can you turn on or off CO
2
or O
2
regulation.
If you attempt to change any of the parameters when the panel is locked, the display will show:
to indicate that a lock is active and needs to be removed before the action can be
performed.
To lock the control panel:
Simultaneously press and hold the SP button and the left arrow for about three seconds:
The display will briefly show: to indicate that the control panel has been locked.
To unlock the control panel:
Simultaneously press and hold the SP button and the right arrow for about three seconds:
The display will briefly show: to indicate that the control panel has been unlocked.