EmbryoScope time-lapse incubator
- 58 -
9.4.1.1 Control unit responses to malfunction information signals
When a warning information signal is activated, the control units will respond as follows:
Response when a
malfunction information
signal is activated
Response when the problem
is solved
A beeping signal will sound for
one second followed by an 11-
second break. The signal is
then repeated.
A flashing blue LED will be
switched on unless the PC is in
error.
The beeping signal will be
switched off.
The flashing blue LED will be
switched off.
PC integrated in the
EmbryoScope time-lapse
incubator
Data about the malfunction
information signal will be saved
in the database of the
EmbryoScope time-lapse
incubator.
Connected EmbryoViewer
software
Malfunction information signals
will be recorded and logged in
the data files for all
EmbryoSlide culture dishes
unless the signal was activated
because the PC was in error.
Activated malfunction
information signals will be
communicated to the external
alarm system (see also section
9.5).