5.3 Temporarily pausing audible warning alarms
To pause the audible alarm signal for three minutes, press the pause audio button: . Pausing
the warning alarm will not affect the LED, which will continue to flash until the problem that caused
the warning alarm has been solved.
5.4 Incubator temperature
5.4.1 Changing the temperature set point
1. In the set point menu, make sure that the display shows the incubator temperature. If this is
not the case, navigate to the temperature menu item by pressing the right or left arrow
until the current temperature set point is displayed: . (temperature menu
item the set point is: 37°C).
2. Press the button. The display will start to alternate between showing .
(temperature menu item the set point is: 37°C) and . (set point menu is active
the set point is: 37°C). This indicates that you are about to adjust the temperature set
point.
3. While still holding the button, press the up or down arrow to increase or
decrease the set point. Each time you press the up or down arrow once, you will either
increase or decrease the set point by 0.1°C.
4. Release the button to save the new setting.
5.4.2 Validating the temperature
The EmbryoScope time-lapse incubator must have been turned on for at least three hours before
the temperature is validated. The room temperature must be equivalent to normal laboratory
temperature. The temperature validation is performed by measuring the temperature in the culture
dish holder using a temperature probe connected to a high-precision thermometer.
NOTE
While performing this procedure, be very careful not to inadvertently touch or push the
EmbryoSlide culture dish(es) present in the culture dish holder.
Before you start, make sure that the high-precision thermometer is calibrated according
to the manufacturer’s specifications.