• Disinfection of the EmbryoScope time-lapse incubator involves unmounting the culture
dish holder. Always perform a camera reset when the culture dish holder has been
unmounted.
Wearing gloves and using good handling techniques are important to successful cleaning.
Follow the below procedure to disinfect the EmbryoScope time-lapse incubator in case of
contamination and/or spillage. The first part of the procedure involves unmounting the culture dish
holder:
1. First end all running EmbryoSlide culture dishes by ending each culture dish individually.
2. Check on the home screen that all culture dishes have been ended.
3. Close the PC by pressing Shutdown on the home screen.
4. Turn off the EmbryoScope time-lapse incubator by pressing the green main switch found
at the rear panel.
5. Open the load door to the embryo chamber.
6. Remove the plate that covers the inaccessible positions of the culture dish holder.