EmbryoScope time-lapse incubator
- 56 -
9.2.2 Operator response to an incubator malfunction alarm
If an incubator malfunction alarm has been activated, turn off the EmbryoScope time-lapse
incubator using the main switch located at the back of the cabinet. Then remove all EmbryoSlide
culture dishes by following the emergency procedure described in section 10.
9.3 Technical alarms
A technical alarm will be activated if the power supply to the EmbryoScope time-lapse incubator is
completely cut off.
9.3.1 Control unit responses to technical alarms
When a technical alarm has been activated, the control units will respond as follows:
The event will immediately be
communicated to the external
alarm system without any delay
(see section 9.5.1).
9.3.2 Operator response to technical alarms
If there is a complete power failure, turn off the EmbryoScope time-lapse incubator using the main
switch located at the back of the cabinet. Then remove all EmbryoSlide culture dishes by following
the emergency procedure described in section 10.