EmbryoScope time-lapse incubator
- 59 -
9.4.2 Warning information signals
The EmbryoScope time-lapse incubator will issue a warning information signal if:
The integrated PC has been in error for more than five minutes.
The software integrated in the EmbryoScope time-lapse incubator is not running.
The connection between the incubator and the integrated PC has been interrupted
(possibly because the integrated PC is shut down).
The load door to the embryo chamber has been left open. The information signal will be
activated after 30 seconds.
The EmbryoScope time-lapse incubator has been paused for more than one hour without
reinsertion of an EmbryoSlide culture dish.
The EmbryoScope time-lapse incubator has been left in Check mode for more than half an
hour.