EmbryoScope time-lapse incubator
- 77 -
14 Planned service and maintenance
A person certified by Vitrolife will inspect and replace all of the following items in accordance with
the intervals specified in the below table.
Replaceable item
Description
Service interval
(years)
Replaced by
VOC HEPA filter
VOC HEPA filter
0.5
Certified service
personnel
O
2
sensor
1 x sensor placed on
the mixing chamber
0.5
Certified service
personnel
UV light
Internal UV light
placed in the mixing
chamber
0.5*
1**
Certified service
personnel
Internal blower
Primary fan
5
Certified service
personnel
External HEPA filters
External HEPA filters
on the gas inlet side
3
Certified service
personnel
Internal mini HEPA
filters
Internal HEPA filters
placed just inside the
gas entrance
5
Certified service
personnel
Proportional valves
Internal valves placed
on the gas system
6
Certified service
personnel
12 V power supply
unit
Internal 12 V power
supply unit
5.5
Certified service
personnel
The UV light service interval differs between ballasts, which is indicated by * (UV light ballast,
220 V) and ** (HF-Matchbox Blue 109 LH-PLS).
To ensure error-free operation, regular service checks are mandatory. We recommend that the
service check be performed by a person certified by Vitrolife.
The default frequency for scheduled periodic service checks is six months.