9.1.3 Multiple warning alarms
If multiple warning alarms occur at the same time, the display on the control panel will show the
warning alarms in order of priority:
The O
2
alarm (activated when the O
2
concentration exceeds 24%) has the highest priority.
Therefore, if an O
2
alarm is activated, this will be the only alarm shown in the display until the
problem has been solved, even if other warning alarms have occurred as well.
The temperature warning alarm has the second highest priority. This alarm will overrule any other
alarm occurring at the same time (except for the O
2
alarm) until the problem has been solved.
If other gas concentration and pressure alarms occur simultaneously, the display will alternate
between these alarms. As the individual problems are solved, they will no longer be shown in the
display.
9.2 Incubator malfunction alarms
An incubator malfunction alarm will be activated if there is a malfunction or power failure of the unit
that controls the running conditions of the EmbryoScope time-lapse incubator.