EmbryoScope time-lapse incubator
- 61 -
9.5 External alarm system
The external alarm system will only be activated when a number of pre-defined errors occur.
Below, you find a list of the errors that will activate the external alarm.
Errors that are not included on this list will not set off the external alarm.
Warning alarms:
Temperature alarms
CO
2
concentration alarms
CO
2
pressure alarms
O
2
concentration alarms
N
2
pressure alarms.
Incubator malfunction alarm:
There is an incubator malfunction or power failure of the unit that controls the running
conditions of the EmbryoScope time-lapse incubator.
Technical alarm:
The power to the entire EmbryoScope time-lapse incubator is completely cut off.
Malfunction information signals:
Motor errors, e.g. if the motors are incorrectly positioned
Camera errors
The user has failed to respond to a displayed dialogue box for more than 15 minutes (the
time delay can be defined by the user)
An error occurred while initialising the Hall sensor.
See also sections 9.1.1, 9.2.1, 9.3.1 and 9.4.2.1 for an in-depth description of which alarms and
information signals will set off the external alarm.