EmbryoScope time-lapse incubator
- 48 -
7. Place the motor lock in lock position. The motor lock is the red device that locks the X and
Y motors.
8. Remove the culture dish holder by loosening the two bolts that keep the holder in place.
An Allen key can be found together with the emergency procedure beneath the service lid.
9. Pull the culture dish holder gently towards you by using the Allen key.
10. Clean all internal surfaces plus the cover plate and culture dish holder (which has been
unmounted from the EmbryoScope time-lapse incubator). Apply EcaFlo Anolyte to sterile
wipes and wipe all internal surfaces and the culture dish holder using at least three wipes.
Repeat until the wipes are no longer discoloured.
11. Change gloves, and after 10 minutes of contact time, spray sterile water over all surfaces
and wipe with a sterile polyester wipe. Alternatively, wipe with a polyester wipe dampened
with sterile water.
12. Repeat steps 10 and 11 three times.
Motor lock in lock