EmbryoScope time-lapse incubator
- 30 -
The Start button will be active.
4. Press the Start button.
5. After three hours, validate the temperature and gas levels using your thermometer and gas
analyser (see also sections 5.4.2, 5.5.2 and 5.6.2).
When you have confirmed that temperature and gas are at the desired levels, the
EmbryoScope time-lapse incubator is ready for insertion of an EmbryoSlide culture dish
(see also section 6.3).
WARNING
During initial start-up and after an incubator shutdown, always validate the gas and
temperature levels using calibrated external validation devices as detailed in the
EmbryoScope user manual. Do NOT rely solely on values displayed on the incubator
control panel.