EmbryoScope time-lapse incubator
- 32 -
8. When the image of well no. 1 is displayed, press the centre of the well on the touch screen.
NOTE
Camera calibration should always start with an image of well no. 1 in the EmbryoSlide
culture dish. Check in the left pane that well no. 1 is highlighted. If this is not the case, a
correct calibration cannot be performed.
Well no. 1 of the culture dish in position no. 1