EmbryoScope time-lapse incubator
- 70 -
When received, all shipping boxes should be inspected for signs of damage during transportation.
In case the boxes are damaged, please contact Vitrolife immediately for further instructions. Do
NOT open the boxes. Leave the EmbryoScope time-lapse incubator in the shipping boxes in a dry
and safe place until it can be handled by a person certified by Vitrolife.
Behaviour in case EMC immunity test levels are exceeded:
If the incubator is subjected to EMC immunity levels exceeding the levels tested, malfunctions and
instabilities may occur, e.g. alarms and screen flickering.
Fuses:
ID
Breaking
capacity
Operating speed
and current
Temperature
Voltage (AC)
Component
F1
100 A (L)
T2,5 A
N/A
250 V
Incubator
F2
100 A (L)
T1,6 A
N/A
250 V
Motor controller
F3
100 A (L)
T2,5 A
N/A
250 V
Integrated PC
F4
100 A (L)
T1,6 A
N/A
250 V
Touch screen
F6
8 A (L)
10 A
72C
250 V
Temperature
F7
100 A (L)
T1,6 A
N/A
250 V
Electronics
F230 VAC
35 A (L)
2 x T3,15 A
N/A
250 V
EmbryoScope time-
lapse incubator