EasyManua.ls Logo

Waters 2707 - Page 76

Waters 2707
100 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-6 Troubleshooting
If no injection occurs:
Cause Solution
Blockage in flow path 1. Disconnect the needle from the valve.
2. Start a manual wash.
3. If solvent flows from the injection port, inspect
the needle; if no solvent flows from the
injection port, disconnect the buffer tubing
from the injector valve.
4. Start a manual wash.
5. If solvent flows from the open end, inspect the
rotor seal; if not, disconnect the buffer tubing
from the syringe valve.
6. Start a manual wash.
7. If solvent flows from syringe valve, inspect the
buffer tubing; if not, check for overtightened
connections throughout the flow path and
check the syringe valve.
Leakage in the injection
valve
1. Disconnect the needle tubing and buffer
tubing.
2. Connect port 1 to an HPLC pump.
3. Block port 6.
4. Start the pump at a low flow.
5. Observe ports 3 and 4 for leakage.
6. If leakage occurs at ports 3 and 4, inspect the
rotor seal; if not, reinspect with a manual
valve.
Tip: Observe the maximum allowed pressure of
350 psi to prevent valve leakage.

Table of Contents

Other manuals for Waters 2707

Related product manuals