WM 67851 01/2021
MEDUVENT Standard EN 101
8 User menu
Result The ventilation parameters of the set ventilation mode are adapted
to match the patient group selected.
8.3.2 Alarm limits
Risk of injury due to alarm limits which are too high or too
low!
Alarm limits which are either too high or too low can prevent the
device from emitting an alarm, thereby putting the patient at risk.
Always set alarm limits which have been adapted to the
patient.
Alarm limits submenu
Alarm Setting range
Activate
automatic
alarm limits
The device automatically sets the respiratory physiology
alarm limits. The automatic limit value is 130 % of the
measured value at the point of activation. At the start of
ventilation, the deviation is 130 % of the value set for the
ventilation parameter.
f 1/min to 140/min