70 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
7 Hygienic reprocessing
The following sections set out the procedures necessary for
hygienic reprocessing. The chapter is divided into the following
sections:
Device
Accessories and other parts
Breathing circuit
Read this chapter in full before starting hygienic reprocessing. If
you have any questions regarding hygienic reprocessing, contact
the manufacturer, WEINMANN Emergency, or a technician
specifically authorized by WEINMANN Emergency.
Disrupted and failed therapy due to incorrect use of
disposables!
Reusing and reprocessing disposables might induce unpredictable
reactions as a result of aging, embrittlement, wear, thermal
loading, and chemical action. This might place the functionality
and safety of the device at risk, and cause the patient or user
serious or life-threatening injury.
Do not reuse disposables.
Do not subject disposables to hygienic reprocessing.
Infection of the user or of the next patient resulting from
incorrect handling of a contaminated hygiene filter!
A contaminated hygiene filter might cause the patient or user
serious or life-threatening injury.
Always wear suitable personal protective equipment when
removing a contaminated hygiene filter.
Dispose of a contaminated hygiene filter when carrying out
hygienic reprocessing, and do not reuse it.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals