EasyManua.ls Logo

Weinmann MEDUVENT Standard - 10 Description of the modes; 10.1 IPPV

Weinmann MEDUVENT Standard
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WM 67851 01/2021
MEDUVENT Standard EN 121
10 Description of the modes
10 Description of the modes
10.1 IPPV
IPPV mode is used for mandatory, volume-controlled ventilation at
a fixed tidal volume. This mode is used on patients who have no
spontaneous breathing. However, a spontaneously breathing
patient can breathe deeply and freely during expiration. The set
maximum pressure limit (pMax) ensures the safety of the patient.
Description
Abbreviation IPPV
Long form Intermittent Positive Pressure Ventilation
Type Volume-controlled
Requirements None
Ventilation parameters
pMax
•Freq.
•Vt
•PEEP
Druck
Zei
t
IPPV
pMax
1/Freq. (eingestellt)
maschinelle Beatmung maschinelle Beatmung
PEEP
1/Freq. (eingestellt)
IPPV
PEEP
1/freq. (set)
Pressure
pMax
1/freq. (set)
Mechanical ventilation
Mechanical ventilation
Time

Table of Contents