74 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
7.2 Preparing for hygienic reprocessing
Requirements The device is switched off (see “6.8 Switching off the device”,
page 69).
The device is disconnected from the patient.
1. Disconnect the device from the power supply.
2. Remove accessories from the device.
3. Disassemble the reusable breathing circuit into its constituent
parts (see “7.3.1 Disassembling the reusable breathing
circuit”, page 74).
4. If necessary: Disassemble the accessories into their constituent
parts.
5. Dispose of all disposables properly (see “14 Disposal”, page
149).
Result All parts have been prepared for hygienic reprocessing.
7.3 Disassembling/assembling the
reusable breathing circuit
The illustrations in this section show all the possible parts of the
reusable breathing circuit. Depending on type, your reusable
breathing circuit might not include some parts.
7.3.1 Disassembling the reusable breathing circuit
Requirements The device is disconnected from the breathing circuit.
The patient is disconnected from the breathing circuit.
1. Open the protective sleeve.
2. Open the hook and loop fasteners in the protective sleeve.
3. Detach the protective cap from the patient end of the reusable
breathing circuit.
4. Disconnect the connecting cable of the MEDUtrigger from the
patient end of the reusable breathing circuit.