WM 67851 01/2021
MEDUVENT Standard EN 83
7 Hygienic reprocessing
7.5 Disinfecting parts by wiping
Permitted parts
Requirements The parts have been cleaned manually and are visibly clean (see
“7.4 Cleaning parts manually”, page 79).
1. For parts approved for disinfection by wiping, refer to the
cleaning and disinfection plan (see “7.11 Cleaning and
disinfection plan”, page 92).
2. For the cleaning agents, dose and exposure time for the
individual parts, refer to the cleaning and disinfection plan (see
“7.11 Cleaning and disinfection plan”, page 92).
3. Prepare the solution for disinfection as specified by the
disinfectant manufacturer.
4. Disinfect parts by wiping with one of the products listed (see
“7.11 Cleaning and disinfection plan”, page 92).
Please note:
Wet uneven surfaces and grooves (e.g. navigation knob,
ventilation hose connection) adequately with disinfectant.
When replacing the filter: Disinfect the filter compartment
by wiping.
Part Disinfection by wiping
Device
Wipe down with Incidin™ Oxywipe S (Ecolab)
Testing bag
12 V cable
Charging adapter
Filter compartment
When replacing the filter:
Wipe down with Incidin™ Oxywipe S (Ecolab)