WM 67851 01/2021
MEDUVENT Standard EN 43
4 Preparation
4.5 Converting the device
4.5.1 Converting the device to a disposable
breathing circuit
1. Remove the protective cap for the connection terminal from
the bore on the measuring circuit connector.
2. Fit the disposable breathing circuit adapter on the measuring
circuit connector.
3. Secure the disposable breathing circuit adapter using the screw
supplied.
4. Fit the protective cap for the connection terminal in the
disposable breathing circuit adapter to cover the screw.
Result The device has been converted for use with a disposable breathing
circuit. The disposable breathing circuit adapter remains on the
device.