44 EN MEDUVENT Standard
WM 67851 01/2021
4 Preparation
4.5.2 Converting the device to a reusable breathing
circuit
1. Release the protective cap for the connection terminal from the
disposable breathing circuit adapter.
2. Loosen the screw on the disposable breathing circuit adapter.
3. Remove the disposable breathing circuit adapter from the
device.
4. Fit the protective cap for the connection terminal on the open
bore on the measuring circuit connection.
Result The device has been converted for use with a reusable breathing
circuit.