16 Technical data
16 Technical data
16.1 Device
Specification Device
Product class according to Directive
93/42/EEC
IIb
Dimensions (W x H x D) 206 mm x 137 mm x 130 mm
Weight:
Without rechargeable battery
With rechargeable battery
1750 g
2100 g
Center of gravity
X = 87 mm
Y = 69 mm
Z = 54 mm
Operation:
Temperature range
Temperature range for oxygen inlet
tube
Humidity
Air pressure
Altitude above mean sea level
-20 °C to +50 °C
-18 °C to +50 °C
5 % to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
Storage (device)/transport:
Temperature range up to 48 h
Temperature range longer than 48 h
Temperature range for oxygen inlet
tube
Humidity
Air pressure
Altitude above mean sea level
-40 °C to +70 °C
-20 °C to +40 °C (recommended: 0 °C to +25 °C)
-18 °C to +50 °C
15 % rh to 95 % rh, no condensation
540 hPa to 1100 hPa
-500 m to 5000 m
Warm-up time from minimum storage
temperature to standby at 20 °C
8 hours
Cooling-down time from maximum
storage temperature to standby at
20 °C
8 hours
Electrical rating
• Rated voltage: 12 V
• Permitted input voltage range: 10 V to 16 V
Max. power consumption 60 W
Current consumption 0.15 A to 4 A