90 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
7.9 Steam-sterilizing parts (optional)
If you intend to perform steam sterilization, do so in accordance
with your in-house procedures.
Requirements The parts intended for steam sterilization are visibly clean.
The parts intended for steam sterilization are disinfected.
1. For parts approved for steam sterilization, refer to the cleaning
and disinfection plan (see “7.11 Cleaning and disinfection
plan”, page 92).
2. Steam-sterilize parts using a device conforming to EN 285.
Please note:
Use a temperature of 134 °C and a dwell time of 5 minutes
or
Use a temperature of 132 °C and a dwell time of 4 minutes
The instructions of the steam sterilizer manufacturer must
be followed.
Result The parts have been steam-sterilized.