7 Hygienic reprocessing
7.10 Preparing parts for reuse
Requirements The parts have been subjected to hygienic reprocessing according
to the cleaning and disinfection plan.
1. Check all parts for damage resulting from use (e.g. tension
cracks or cable breaks).
2. Replace damaged parts.
3. Assemble the reusable breathing circuit (see
“7.3.2 Assembling the reusable breathing circuit”, page 77).
4. Install the accessories.
5. Reconnect the power supply (see “4.2 Connecting a power
supply”, page 34).
6. Perform a function check (see “5 Function check”, page 45).
7. Place parts in storage in accordance with the storage
conditions (see “16 Technical data”, page 157).
Result The parts are ready for use again.