122 EN MEDUVENT Standard
WM 67851 01/2021
10 Description of the modes
10.2 CPAP
The CPAP/PEEP setting increases the lower pressure level during
spontaneous breathing. This results in an increase in the functional
residual capacity of spontaneously breathing patients. The patient
is able to breathe spontaneously without any restriction at the set
pressure level. Pressure is generally set at the end of expiration
(PEEP).
Hazardous therapy as a result of an unsuitable ventilation
mode!
In ventilation modes CPAP and CPAP + PS, patients do not receive
any mandatory mechanical breaths. In patients with inadequate
spontaneous breathing or respiratory arrest, use of these modes
might lead to hypoventilation, and cause the patient serious or
life-threatening injury.
Use CPAP or CPAP + PS modes only on patients with adequate
spontaneous breathing.
Continuously monitor the patient and device during ventilation.
Description
Abbreviation CPAP
Long form Continuous Positive Airway Pressure
Type Pressure-controlled
Requirements None
Ventilation parameters
PEEP
CPAP
Druck
Zeit
1/Freq. (spontan) 1/Freq. (spontan)
CPAP / PEEP
CPAP
Pressure
CPAP / PEEP
Time
1/freq. (spontaneous)
1/freq. (spontaneous)

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals