84 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
5. Allow the parts to dry completely.
6. Check the parts for residues and residual soiling.
7. If visible soiling remains: Repeat disinfection by wiping.
Result The parts have been disinfected by wiping.
7.6 Disinfecting parts by immersion
Permitted parts
Requirements The parts intended for disinfection by immersion have been
cleaned manually (see “7.11 Cleaning and disinfection plan”,
page 92). Disinfection of the reusable measuring circuit by means
of immersion is described separately (see “7.7 Disinfecting the
reusable measuring circuit by immersion”, page 86).
1. For parts approved for disinfection by immersion, refer to the
cleaning and disinfection plan (see “7.11 Cleaning and
disinfection plan”, page 92).
Depending on the disinfectant, it might be necessary to wipe over
with a neutralizing product afterward.
Part Disinfection by immersion
Reusable breathing circuit
Ventilation hose
Immerse in gigasept
®
FF new
(Schülke)
Dose: 50 ml/l
Exposure time: 15 min.
Patient valve
Elbow
Protective cap
Service label
Protective sleeve
Reusable measuring circuit, comprising:
• PEEP control tube
• Pressure measuring tube
• Flexible oxygen tube
• Measuring circuit connector
(see “7.7 Disinfecting the reusable
measuring circuit by immersion”, page
86)
Hook and loop strap with clip