10 Description of the modes
10.4 PRVC + PS (optional)
Risk of injury from autotriggering!
Automatic triggering of the inspiratory trigger by artifacts
(autotriggers) can lead to the patient hyperventilating.
Reduce the sensitivity of the inspiratory trigger in the event of
autotriggers.
Description
Abbreviation PRVC + PS
Long form
Pressure-Regulated Volume-Controlled Ventilation
+ Pressure Support
Type Pressure-controlled
Requirements PRVC + PS option has been activated
Ventilation parameters
• pMax
•Freq.
•Vt
•InTr
• ΔpPS
•PEEP
Druck
Zeit
1/Freq. (eingestellt) 1/Freq. (eingestellt) 1/Freq. (eingestellt) 1/Freq. (eingestellt)1/Freq. (spontan)
druck-
kontrollierter
Beatmungs-
hub
druck-
kontrollierter
Beatmungs-
hub
druck-
kontrollierter
Beatmungs-
hub
druck-
kontrollierter
Beatmungs-
hub
PS
pInsp (variabel)
pInsp (variabel)
pInsp (variabel)
PS
pInsp (variabel)
PEEP
pMax
Pressure
Time
pMax
1/freq. (spontaneous)
1/freq. (set)
1/freq. (set) 1/freq. (set)1/freq. (set)
Pressure-
controlled
mechanical
breath
Pressure-
controlled
mechanical
breath
Pressure-
controlled
mechanical
breath
PSPressure-
controlled
mechanical
breath
PEEP
pInsp (variable)
pInsp (variable)
pInsp (variable)
PS
pInsp (variable)