EasyManua.ls Logo

Weinmann MEDUVENT Standard - 16.9 Alarm delay times

Weinmann MEDUVENT Standard
188 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
176 EN MEDUVENT Standard
WM 67851 01/2021
16 Technical data
16.9 Alarm delay times
Alarm Delay time
Apnea 30 s
Vt not achievable
At 5/min
At 40/min
Triggers after 2 breaths
Up to 24 s
3 s
Respiratory rate q
At 5/min
At 40/min
Triggers after 2 breaths
Up to 24 s
3 s
Vt r
At 5/min
At 40/min
Triggers after 2 breaths
Up to 24 s
3 s
PEEP high q
At 5/min to 8/min
At 40/min
Triggers after 2 breaths
5 s
3 s

Table of Contents