WM 67851 01/2021
MEDUVENT Standard EN 175
16 Technical data
result
Overall result of the function check:
• ok = check passed
• failed = check failed
• not tested = check not performed
A function check is considered failed if at
least one test is failed.
alarmsystem
Check of visual and audio alarms including
alarm for Airway pressure high q
• energy failure
buttontest Check of buttons and navigation knob
filterwear Check of hygiene filter
medutrigger Check of MEDUtrigger
powerelectronics Check of electronics
blower Check of blower
flowout
Check of zero point of flowout differential
pressure sensor
flowo2
Check of zero point of flowO
2
differential
pressure sensor
presplausible Check of airway pressure sensor (pneumatic)
expvalvecontrol Check of patient valve control
hosesystemtight Check of breathing circuit for leaks
expvalvetight Check of patient valve for leaks
volplausible Check of volume administered
checkvalvetight
Check of check valve diaphragm in patient
valve
flowoutplausible
Cross-check of flow measurement sensors
(flowout and flowo2)
pawaccurate Check of airway pressure sensor (pneumatic)
deviceconfig Check of device configuration
oxygeninletopen Check of oxygen inlet