86 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
7.7 Disinfecting the reusable measuring
circuit by immersion
The principle described applies to the following parts of the
reusable measuring circuit:
Pressure measuring tube
PEEP control tube
Flexible oxygen tube
Measuring circuit connector
Requirements The reusable measuring circuit has been disconnected from the
reusable breathing circuit (see “7.3.1 Disassembling the
reusable breathing circuit”, page 74).
The reusable measuring circuit has been cleaned manually.
1. For the disinfectants, dose and exposure time for the individual
parts, refer to the cleaning and disinfection plan (see
“7.11 Cleaning and disinfection plan”, page 92).
2. Prepare the solution for disinfection by immersion as specified
by the disinfectant manufacturer.
3. Connect a sterile disposable syringe (20 ml) to a free end of a
tube.
4. Draw the solution for disinfection by immersion through the
tube into the disposable syringe until both are completely full.
5. Disconnect the disposable syringe from the tube.
Risk of infection and contamination resulting from defective
hygienic reprocessing of the measuring circuit!
Flushing the measuring circuit in alternate directions does not
ensure absence of germs, and might injure the patient.
Flush measuring tubes only in one direction.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals