WM 67851 01/2021
MEDUVENT Standard EN 85
7 Hygienic reprocessing
2. For the disinfectants, dose and exposure time for the individual
parts, refer to the cleaning and disinfection plan (see
“7.11 Cleaning and disinfection plan”, page 92).
3. Prepare the solution for disinfection by immersion as specified
by the disinfectant manufacturer.
4. Immerse parts in the solution for disinfection by immersion.
Please note:
• All cavities must be filled.
• There must be no air bubbles.
• All surfaces must be wetted.
• Swirl parts around in the solution for disinfection by
immersion so as to wet all surfaces and any cavities
completely.
• Observe the exposure time specified in the cleaning and
disinfection plan.
5. After the specified exposure time, rinse the parts in water of
potable quality for 5 minutes to remove all disinfectant
residues.
6. Allow the parts to dry completely.
7. Check the parts for residues and residual soiling.
8. In case of visible soiling: Repeat cleaning and disinfection.
Result The parts have been disinfected by immersion.