WM 67851 01/2021
MEDUVENT Standard EN 131
11 Alarms and faults
11 Alarms and faults
Alarms are shown in the alarm line of the display in the form of
text. The text is displayed with a particular background color as a
function of alarm priority:
If more than one alarm is active, the device responds as follows:
Multiple alarms of different priorities: The device displays the
alarm with the highest priority. Alarms with a lower priority do
not appear until the higher-priority alarm is no longer active.
Multiple alarms of identical priority: The device displays the
alarms in rotation.
Technical alarms predominate. They cannot be muted.
Technical alarms are generated if no ventilation by the device
is possible.
Alarm color Priority Meaning
Red High priority
High-priority alarms warn of imminent
fatal or irreversible patient injuries or
of device faults.
Yellow Medium priority
Medium-priority alarms warn of
immediate reversible patient injuries
or of minor device faults.
Turquoise Low priority
Low-priority alarms warn of delayed
minor injuries or inconvenience to the
patient or minor restrictions on the
device.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals