7 Hygienic reprocessing
6. Immerse the tube in the solution for disinfection.
Please note:
• All surfaces and lumina must be wetted completely.
• Observe the exposure time specified in the cleaning and
disinfection plan.
7. Once the exposure time has elapsed: Use the syringe to rinse
the tube at least 8 times in water of potable quality.
Please note: Flush only in one direction.
8. Follow this principle and repeat the procedure for each tube.
9. Allow the tubes to dry completely.
10. If necessary: Dry the tubes with medical compressed air or
medical oxygen.
11. Check the tubes for residues and residual soiling.
12. If visible soiling is present: Repeat disinfection by immersion.
Result The reusable measuring circuit has been disinfected by immersion.