EasyManuals Logo

Weinmann MEDUVENT Standard User Manual

Weinmann MEDUVENT Standard
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #87 background imageLoading...
Page #87 background image
WM 67851 01/2021
MEDUVENT Standard EN 87
7 Hygienic reprocessing
6. Immerse the tube in the solution for disinfection.
Please note:
All surfaces and lumina must be wetted completely.
Observe the exposure time specified in the cleaning and
disinfection plan.
7. Once the exposure time has elapsed: Use the syringe to rinse
the tube at least 8 times in water of potable quality.
Please note: Flush only in one direction.
8. Follow this principle and repeat the procedure for each tube.
9. Allow the tubes to dry completely.
10. If necessary: Dry the tubes with medical compressed air or
medical oxygen.
11. Check the tubes for residues and residual soiling.
12. If visible soiling is present: Repeat disinfection by immersion.
Result The reusable measuring circuit has been disinfected by immersion.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals