WM 67851 01/2021
MEDUVENT Standard EN 77
7 Hygienic reprocessing
7.3.2 Assembling the reusable breathing circuit
Requirements The reusable breathing circuit has been disassembled.
1. Attach the band of the protective cap to the check valve
diaphragm holder.
2. Fit the patient valve.
Please note:
The side of the PEEP control diaphragm marked "TOP"
must be facing up toward the control cover.
The arrow on the control cover must be pointing toward
the patient.