82 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
6. Use the disposable syringe to fill the pressure measuring tube
and the flexible oxygen tube with cleaning solution in the same
way.
7. Immerse the reusable measuring circuit in cleaning solution.
Please note:
All surfaces and lumina must be wetted completely.
Observe the exposure time specified in the cleaning and
disinfection plan.
8. Rinse the outside of the reusable measuring circuit with water
of potable quality.
9. Using the disposable syringe, rinse the inside of the reusable
measuring circuit at least 8 times with water of potable quality.
Please note: Flush only in one direction.
10. Allow the reusable measuring circuit to dry completely.
11. If necessary: Allow the ventilation hose to dry completely.
12. Check the reusable measuring circuit for residues and residual
soiling.
13. If visible soiling is still present: Repeat manual cleaning.
Result The reusable measuring circuit has been cleaned manually.