WM 67851 01/2021
MEDUVENT Standard EN 71
7 Hygienic reprocessing
Disrupted or failed therapy due to unsuitable cleaning agents
and disinfectants!
Using incorrect cleaning agents and disinfectants might cause the
device to malfunction. This can cause serious or life-threatening
injury to the patient.
Never clean the device and accessories with bleach, bleach
solution, or compounds containing phenols.
Use only the cleaning agents and disinfectants recommended
in these instructions for use (see “7.11 Cleaning and
disinfection plan”, page 92).
Loss of mechanical or electrical safety resulting from
reprocessing of the device and accessories with unsuitable
cleaning agents and disinfectants!
Using incorrect cleaning agents and disinfectants might cause
damage to the surface of the device and accessories, as well as
impairing electrical and insulating properties. This might cause the
user and the patient serious or life-threatening injury.
Use only the cleaning agents and disinfectants recommended
in these instructions for use (see “7.11 Cleaning and
disinfection plan”, page 92).
Disrupted or failed therapy due to liquid in the breathing
circuit after hygienic reprocessing of the reusable measuring
circuit!
Droplets inside the reusable measuring circuit might falsify the
results it delivers. This can cause serious or life-threatening injury
to the patient.
After hygienic reprocessing of the reusable measuring circuit,
allow all the tubes in the breathing circuit to dry thoroughly.