Page #128 background imageLoading...
Page #128 background image
128 EN MEDUVENT Standard
WM 67851 01/2021
10 Description of the modes
Risk of intrinsic PEEP!
If the expiration phase is too short, pressure might rise slowly at
the end of it. This can cause serious or life-threatening injury to
the patient.
Set the pressure limit correctly.
Monitor the patient continuously.
Description
Abbreviation S-IPPV
Long form
Synchronized Intermittent Positive Pressure
Ventilation
Type Volume-controlled
Requirements S-IPPV option has been activated
Ventilation parameters
pMax
•Freq.
•Vt
•InTr
•PEEP
Druck
Zeit
S-IPPV
pMax
1/Freq. (eingestellt)
maschinelle Beatmung
synchronisierte maschinelle
Beatmung
1/Freq. (aktuell)
Triggerfenster
ΔT
P
Plateau
PEEP
S-IPPV
Pressure
pMax
Mechanical ventilation
Time
Trigger window
1/freq. (current)
PEEP
Synchronized mechanical
ventilation
1/freq. (set)
P
plateau

Table of Contents