11 Alarms and faults
11.1.2 Medium-priority alarms (yellow)
Alarm Cause Remedy
Battery defective
Battery defective
Run device on rechargeable
battery without line power until it
switches off. Fully recharge
battery. If the device continues to
display the alarm: Replace
rechargeable battery.
Rechargeable battery not inserted,
or not inserted correctly
Insert rechargeable battery
correctly.
Battery weak
Rechargeable battery state of
charge low
Connect device to line power and
charge rechargeable battery.
Check battery
Wrong rechargeable battery
inserted
Insert approved rechargeable
battery.
Frequency high q
Patient's respiratory rate too high
Check the patient's condition.
Check limit value settings for
plausibility.
Inspiratory trigger triggered by
artifacts (autotriggers)
Check mask for leaks.
Reduce the sensitivity of the
inspiratory trigger.
Oxygen inlet flow higher than
necessary
Flow setting higher than necessary
Reduce flow setting in steps. Rule
of thumb for 100 % oxygen: Flow
= MVi.
Oxygen inlet leakage
Oxygen inlet is not sealed and no
oxygen is being supplied.
Seal oxygen inlet with protective
cap for oxygen inlet or supply
oxygen.
Service required Device defective Have the device repaired.
Vt not achievable
Implausible ventilation parameters Adjust ventilation parameters.
pMax set too low Modify setting for pMax.