12 Maintenance
12.4.1 Removing the hygiene filter
1. Press together and hold the tabs of the locking mechanism.
2. Pull the hygiene filter out of the device's filter compartment.
3. Disinfect the filter compartment by wiping.
Result The hygiene filter has been removed.
12.4.2 Inserting the hygiene filter
Infection of the user or of the next patient resulting from
incorrect handling of a contaminated hygiene filter!
A contaminated hygiene filter might cause the patient or user
serious or life-threatening injury.
Always wear suitable personal protective equipment when
removing a contaminated hygiene filter.
Dispose of a contaminated hygiene filter when carrying out
hygienic reprocessing, and do not reuse it.
Risk of contamination or infection resulting from impaired
filter properties!
Soiling, foreign bodies or damage in the filter compartment or on
the hygiene filter might mean that the filter element is not
correctly positioned. As a result, the system will not be leak-tight,
and contamination or infection might occur which might cause
the patient or user serious or life-threatening injury.
Check the filter compartment and the hygiene filter for soiling,
foreign bodies and damage.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals