EasyManuals Logo

Weinmann MEDUVENT Standard User Manual

Weinmann MEDUVENT Standard
188 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #16 background imageLoading...
Page #16 background image
16 EN MEDUVENT Standard
WM 67851 01/2021
2 Safety
Disrupted or failed therapy due to magnetic resonance
imaging machines in the immediate vicinity of the device!
The magnetic action of magnetic resonance imaging machines in
the immediate vicinity of the device might throw the device
around. This might lead to therapy being interrupted, and cause
the patient serious or life-threatening injury.
Never operate device in conjunction with magnetic resonance
imaging machines.
Disrupted or failed therapy due to portable radio-frequency
communication equipment in the immediate vicinity of the
device!
Portable high-frequency communication equipment (e.g. radios,
antennas and antenna cables) in the immediate vicinity of the
device may impair the performance characteristics of the device,
and injure the patient.
Keep portable radio-frequency communication equipment a
minimum distance of 30 cm away from the device and its
accessories.
Faults in, or failure of, the device or its accessories during
therapy as a result of high-frequency surgical equipment in
the immediate vicinity of the device!
High-frequency surgical equipment in the immediate vicinity of
the device or its accessories may lead to faults or failure of the
device or its accessories. This can cause serious or life-threatening
injury to the patient.
Do not use the device and its accessories in the vicinity of high-
frequency surgical equipment.
Disrupted or failed therapy due to incompatibility of the
device with consumables, accessories and other medical
devices!
Defective or non-approved accessories might cause malfunctions,
increased electromagnetic interference or reduced
electromagnetic immunity of the device, incorrect output values,
and reduced ventilation performance. This can cause serious or
life-threatening injury to the patient.
Use only approved accessories.
Follow the instructions for use of the accessories.

Table of Contents

Other manuals for Weinmann MEDUVENT Standard

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUVENT Standard and is the answer not in the manual?

Weinmann MEDUVENT Standard Specifications

General IconGeneral
BrandWeinmann
ModelMEDUVENT Standard
CategoryMedical Equipment
LanguageEnglish

Related product manuals