16 Technical data
PEEP 0 mbar to 20 mbar (±3 mbar or ±15 %)
Trigger
Can be set to Levels 1 to 3
(not in IPPV, CPAP and Manual modes)
Airway pressure sensor -5 hPa to 80 hPa, measurement position close to patient
Accuracy of airway pressure
measurement
-5 hPa to 80 hPa (±5 % or ±1.5 hPa)
Pressure support ΔpPS (only with CPAP
+ PS and PRVC + PS options)
0 hPa to 30 hPa (±3 hPa or ±15 %) above PEEP
Volume sensor
-30 l/min to 150 l/min, measuring position ventilation hose
connection (BTPS oder ATP, whichever value is lower)
Accuracy of measurement of respiratory
minute volume (MVi)
±20 % or ±1.2 ml/min (BTPS, whichever value is higher)
Accuracy of measurement of tidal
volume (Vti)
±20 % or ±40 ml (BTPS, whichever value is higher)
Gas composition
Mixture of air, oxygen, CO
2
. Oxygen fraction 21 % to 100 %, CO
2
fraction 0 % to 10 %
Ventilation hose connection 22 mm outer cone
Patient valve connections
22 mm outer cone
15 mm inner cone
Hygiene filter service life
Service life: 6 months
Operating time as a function of filter wear:
• Filter wear normal, 100 %: 24 h operation
• Filter wear high, 150 %: 16 h operation
• Filter wear very high, 200 %: 12 h operation
Hygiene filter separation rate > 99 %
Specification Device
0
2
4
6
8
10
12
14
16
0 200 400 600 800 1000 1200 1400 1600 1800 2000
Stufe 3
Stufe 2
Stufe 1
Triggerempfindlichkeit
Vt / ml
Trigger / l/min
Trigger sensitivity
Trigger / l/min
VT / ml
Level 1
Level 2
Level 3