16 Technical data
Medical electrical equipment is subject to special precautions in
relation to electromagnetic compatibility (EMC). It must be
installed and put into operation in accordance with the EMC
information contained in the accompanying documents.
Disrupted or failed therapy due to portable radio-frequency
communication equipment in the immediate vicinity of the
device!
Portable radio-frequency communication equipment (e.g. radios,
antennas and antenna cables) in the immediate vicinity of the
device might affect the functioning of the device. This can cause
serious or life-threatening injury to the patient.
Keep portable radio-frequency communication equipment a
minimum distance of 30 cm away from the device and its
accessories.
Guidelines and manufacturer's declaration - emission of electromagnetic interference
MEDUVENT Standard is designed for operation in the electromagnetic environment described below. The
customer or user of the MEDUVENT Standard device should ensure that it is operated in such an
environment.
Measurements of
interference emission
Compliance Electromagnetic environment - guidelines
RF emissions acc. to
CISPR 11
Group 1,
Class B
The RF emission of MEDUVENT Standard is very low,
and it is unlikely to interfere with adjacent electronic
devices.
Emission of oscillations acc.
to IEC 61000-3-2
Compliant
MEDUVENT Standard is suitable for use in all
facilities, including domestic environments, and
those which are connected directly to the public
power grid which also supplies buildings used for
residential purposes.
Emission of voltage
fluctuations/flicker acc. to
IEC 61000-3-3
Compliant
RF emissions acc. to
RTCA DO-160 G
Section 21,
Category M
MEDUVENT Standard is suitable for use in operating
locations of Category M inside aircraft due to its low
RF emission.
RF emissions acc. to UN/ECE
Rule no. 10
Annex 6, Annex 7
MEDUVENT Standard is suitable for use in motor
vehicles due to its low RF emission.
Emissions interfering with
motor vehicle power supply
lines acc. to ISO 7637-2
Compliant
MEDUVENT Standard is suitable for connection to
the on-board power supply due to its low RF
emission.