16 Technical data
710
704 to 787
LTE Band 13,
17
Pulse modulation
b
217 Hz
0.2 0.3 9745
780
810
800 to 960
GSM 800/900,
TETRA 800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse modulation
b
18 Hz
2 0.3 28
870
930
1720
1700 to 1990
GSM 1800,
CDMA 1900,
GSM 1900
DECT,
LTE Band 1, 3,
4, 25, UMTS
Pulse modulation
b
217 Hz
2 0.3 28
1845
1970
2450 2400 to 2570
Bluetooth,
WLAN 802.11
b/g/n
RFID 2450
LTE Band 7
Pulse modulation
b
217 Hz
2 0.3 28
5240
5100 to 5800
WLAN 802.11
a/n
Pulse modulation
b
217 Hz
0.2 0.3 95500
5785
a
For some radio services, only the frequencies for uplinking the mobile communication equipment to the
base station were included in the table.
b
The carrier must be modulated with a square wave with a 50 % signal ratio.
c
As an alternative to frequency modulation (FM), pulse-width modulation with a 50 % 18 Hz signal ratio
can be used, as this would represent the worst-case scenario even if it is not the actual modulation.
Guidelines and manufacturer's declaration - electromagnetic immunity
MEDUVENT Standard has been tested for immunity to the radio services listed below.
If the field strength measured at the site where MEDUVENT Standard is used exceeds the above compliance
levels, MEDUVENT Standard should be monitored to provide evidence of function in accordance with its
intended purpose. If unusual performance characteristics are observed, additional measures may be
required, such as a different orientation or location for MEDUVENT Standard.
Testing
fre-
quency
Frequency
band
a
Radio
service
a
Modulation
b
Max.
output
power
Dis-
tance
Immunity
test level
MHz MHz W m V/m