6 Operation
Hazardous therapy due to lack of monitoring of the oxygen
concentration administered!
The device does not monitor inspiratory oxygen concentration in
the same way as an RGM (respiratory gas monitor), and has no
corresponding alarm. Dispensing respiratory gas with a differing
oxygen concentration might place the therapy at risk. This might
injure the patient.
Use a separate respiratory gas monitor to monitor the oxygen
concentration administered to the patient.
Inadequate oxygen concentration due to leaks during
ventilation!
Leaks during ventilation may lead to the concentration of oxygen
administered in the respiratory gas being reduced. This might
injure the patient.
Check that the mask or tracheal tube is correctly positioned.
Note measured value O
2
i for assessing ventilation. Also correct
oxygen supply if necessary.
Use suitable external patient monitoring methods.