64 EN MEDUVENT Standard
WM 67851 01/2021
6 Operation
4. Repeat the steps for all the ventilation parameters you want to
change.
Result Ventilation parameters have been changed.
The following parameters are mutually dependent:
pMax/PEEP:
The minimum difference between pMax and PEEP is 5 mbar.
pMax PEEP + 5 mbar
•pMax/ΔpPS:
The minimum difference between PEEP + ΔpPS and pMax is
1 mbar if ΔpPS 5 mbar.
ΔpPS < 5 mbar: pMax PEEP + 5 mbar
ΔpPS 5 mbar: pMax PEEP + ΔpPS + 1 mbar
Frequency/Vti:
The product of Freq. x Vti results in inspiratory minute volume
MVi.
MVi = Freq. × Vti
The frequency and tidal volume ventilation parameters can
only be set in meaningful combinations. Combinations
resulting in a value below 1.5 l/min or above 20 l/min (BTPS)
cannot be set.