80 EN MEDUVENT Standard
WM 67851 01/2021
7 Hygienic reprocessing
1. For parts approved for manual cleaning, refer to the cleaning
and disinfection plan (see “7.11 Cleaning and disinfection
plan”, page 92).
2. For the cleaning agents, dose and exposure time for the
individual parts, refer to the cleaning and disinfection plan (see
“7.11 Cleaning and disinfection plan”, page 92).
3. Prepare the cleaning solution as specified by the cleaning agent
manufacturer.
4. To remove all visible soiling: Brush parts thoroughly inside and
out using a commercially available soft brush suitable for plastic
moistened with the cleaning agent.
Please note:
• Keep uneven surfaces and grooves (e.g. top and bottom of
the MEDUtrigger, navigation knob, ventilation hose
connection) moist throughout exposure time, and brush off
particularly thoroughly.
• Brush hoses/tubes with a special lumen brush.
5. If the cleaning and disinfection plan stipulates that parts have
to be wiped down: Wipe down parts using a clean, lint-free
disposable cloth moistened with cleaning solution.
Please note:
• Use a fresh wipe for every cleaning process.
• Carefully wipe down all surfaces.
• All surfaces must be wetted with cleaning solution.
• Observe the exposure time specified in the cleaning and
disinfection plan.