7 Hygienic reprocessing
7.11.2 Device and accessories
Carry out hygienic reprocessing according to the table below after
every use:
Part
Level of
disinfec-
tion
Manual
cleaning
(only nec-
essary in
case of
visible
soiling)
Disinfec-
tion by
wiping
Disinfec-
tion by
immersion
Mechani-
cal repro-
cessing
Steriliza-
tion
Device
Intermediate-
level
disinfection
Wipe down
with
neodisher
®
MediClean
forte (Dr.
Weigert).
Dose:
10 ml/l
Duration:
Treat all
surfaces at
least twice
until they are
visibly clean
Wipe down
with
Incidin™
Oxywipe S
(Ecolab)
Not permitted Not permitted Not permitted
12 V cable
Charging
adapter
MEDUtrigger
with connect-
ing cable
Filter
compartment
When
replacing the
filter: Wipe
down with
Incidin™
Oxywipe S
(Ecolab)
Hook and loop
strap with clip
Wipe down
with
neodisher
®
MediClean
forte (Dr.
Weigert)
Not permitted
Immerse in
gigasept
®
FF
new (Schülke)
Dose:
50 ml/l
Exposure
time:
15 min.
Wash with
Derval SOLO
and Ottalin
PERACET at
up to 70 °C
Not permitted