7 Hygienic reprocessing
7.11.3 Breathing circuits
Part
Level of
disinfec-
tion
Manual
cleaning
(only nec-
essary in
case of
visible
soiling)
Disinfec-
tion by
wiping
Disinfec-
tion by
immersion
Mechani-
cal repro-
cessing
Steriliza-
tion
Reusable breathing circuit
Ventilation
hose
Intermediate-
level
disinfection
Immerse in
neodisher
®
MediClean
forte (Dr.
Weigert) and
wash.
Dose:
10 ml/l
Duration:
Treat all
surfaces at
least twice
until they are
visibly clean
Not permitted
Immerse in
gigasept
®
FF
new (Schülke)
Dose:
50 ml/l
Exposure
time:
15 min.
Cleaning:
neodisher
®
MediClean
forte (Dr.
Weigert):
0.5 %, 55 °C,
10 min
Thermal
disinfection:
90 °C, 5 min
(corresponds
to A0 value
3000)
Permitted as
an option:
Steam
sterilization*
following
prior disin-
fection:
5 min at
134 °C
or
4 min at
132 °C
Patient valve
Elbow
Protective cap
Reusable mea-
suring circuit